Model Number |
AN9620H(F)/AN9620TH(F) |
|
Model
|
AN9620TH(F)
|
|
Power supply
|
220V ±10%, 50Hz
|
|
Test input Voltage
|
Phase voltage: 30V ~ 300V AC single phase/Three phase
|
|
Test input Current
|
23A AC MAX
|
|
Voltage Display Range/ Resolution/ Accuracy
|
Range: 30.0 ~ 300.0V, Resolution :0.1V, Accuracy:±(1% of reading+0.5V)
|
|
Over Current Protect
|
25A
|
|
Inner MD
|
MD A,C, E, F (default), other MD is optional
|
|
Leakage Current Measurement Range
|
MD A、C、D、E、F: 10uA~6000uA
MD B: 10uA~18.00mA
MD G: 10uA~4500uA
|
|
Leakage Current Measurement Resolution
|
10.0μA~999.9μA Resolution: 0.1μA
1000μA~9999μA Resolution: 1μA
10.00mA~18.00mA Resolution: 10μA
|
|
Accuracy
|
MD:A、B、C、D、E、F
|
10uA~50uA DC~1M ±(2%of reading +3uA)
50uA~MDmax DC~10k ±(2%of reading+0.5uA);
10k~200k ±(5%of reading+0.5uA);
200k~1M ±(5%of reading+1.5uA)
|
|
MD:G
|
10uA~50uA DC~1M ±(2%of reading+12uA)
50uA~100uA DC~1M ±(2%of reading+4uA)
100uA~MDmax DC~10k ±(2%of reading+0.5uA);
10k~200k ±(5%of reading+0.5uA);
200k~1M ±(5%of reading+1.5uA)
|
|
Leakage Current Alarm Preset
|
Upper Limit preset: 0~MAX of measurement range(0=disable)
Lower Limit preset: 0~MAX of measurement range(0=disable)
|
|
Measurement Time Setting Range/ Resolution / Accuracy
|
0~999.9s( 0=continues test); Resolution :0.1s Accuracy: ±(0.1% of setting +0.1s)
|
|
Display Mode
|
LCD
|
|
Interface
|
Remote control(PLC), RS232(RS485 optional)
|
|
Dimension (W×H×D) (mm)
|
400×156×475
|
AN9620TH(F)
Main Features:
◆Original plug and play MD clip,display MD network circuit;
◆Original MD network circuit video,changeable,can be measured;
◆Inner 4 kinds of main MD slip,can equip other slips;
◆DC-1MHz bandwidth,Earth Leakage Current,Enclosure leakage current,and Applied part leakage
◆Can test single phase/three phase test
◆Intelligent judgment,voice and light alarm,large font display;
◆8-group and 8-step test condition memory
Technical Parameters:
|
MD
|
Standard
|
Application Area
|
|
A
|
L3101-1,UL1950,UL60950,IEC60950,IEC60065,IEC61010-1,IEC60335-1,IEC60990,GB/T12113,GB4943,GB8898,GB4706,GB7000
|
Examination and Communication equipments(apperceive and responses current)
|
|
B
|
UL1563,IEC60335-1
|
Electrical equipment and assembly products
|
|
C
|
UL2601,EN60601-1,IEC60601-1,GB9706.1
|
Medical instrument and equipment
|
|
D
|
UL544 Patient Care
|
Medical instrument and equipment
|
|
E
|
UL544 Non Patient, UL484, IEC60598
|
Medical instrument and equipment
|
|
F
|
L3101-1,UL1950,UL60950,IEC60950,IEC60065,IEC61010-1,IEC60335-1,IEC60990,GB/T12113,GB7000
|
Examination and Communication equipments(slip current)
|
|
G
|
IEC60950-1
|
Function Test
|
|